The HomeChoice Pro APD System is a peritoneal dialysis recall
The HomeChoice Pro APD System is a peritoneal dialysis system. The HomeChoice Pro Automated Personal Cycler peritoneal dialysis system is intended for…
Why was The HomeChoice Pro APD System is a peritoneal dialysis recalled?
The occluder in affected devices 1) may have sharp edges that could cut the pneumatic tubing that connects the occluder blade to the manifold causing an unrecoverable system error, and/or 2) blade may separate from the back plate and cause Slow / No Flow alarm requiring therapy termination.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Product Code: 5C8310; Serial Numbers: 80415, 680416, 680418, 680420, 680421, 680422, 680424, 680426, 680427, 680428, 680429, 680430, 680438, 680439, 680440, 680441, 680444, 680445, 680448, 680449, 680452, 680453, 680454, 680455, 680456, 680457, 680459, 680460, 680461, 680462, 680463, 680465, 680467, 680467, 680469, 680470, 680472, 680473, 680475, 680476, 680477, 680478, 680480, 680481, 680482, 680463, 680484, 680485, 680486, 680489, 680490, 680491, 680494, 680495, 680496, 680498, 680499, 680500, 680501, 680502, 680503, 680504, 680505, 680506, 680507, 680508, 680509, 680510, 680516, 680517, 680519, 680520, 680521, 680522, 680524, 680525, 680526, 680527, 680527, 680528, 680529, 680530, 680532, 680533, 680534, 680536, 680537, 680538, 680539, 680541, 680542, 680543, 680547, 680552, 680553, 680544, 680555, 680561, 680562, 680566, 680567, 680585, 680586, 680587, 680588, 680589, 680590, 680591, 680593, 680594, 680595, 680596, 680597, 680598, 680599, 680600, 680601, 680602, 680603, 680604, 680605, 680606, 680607, 680608, 680609, 680611, 680612, 680613, 680614, 680615, 680616, 680617, 680618, 680619, 680620, 680621, 680622, 680623, 680624, 680625, 680626, 680627, 680628, 680629, 680630, 680631, 680632, 680633, 680634
- Product
- The HomeChoice Pro APD System is a peritoneal dialysis system. The HomeChoice Pro Automated Personal Cycler peritoneal dialysis system is intended for…
- Recalling firm
- Baxter Healthcare Corp.
- Quantity
- 75 units (US)
- Distribution
- Worldwide Distribution-US (nationwide) including the states of AK, AR, AZ, CA, FL, GA, HI, IA, ID, IL, IN, KS, LA, MA, MD, MI, MO, NC, NJ, NV, NY, OH, PA, SC, TN, TX, UT, VA, VT, WA, and WI, and the countries of Australia, Canada, and Mexico.
- Recall date
- March 25, 2015
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)