FDAClass IMarch 20, 2024

Cerenovus Cerebase DA Guide Sheath recall

CERENOVUS CEREBASE DA Guide Sheath, Part Numbers: a) GS9080SD; b) GS9090SD; c) GS9095SD; Vascular guide-catheter, single-use

Why was Cerenovus Cerebase DA Guide Sheath recalled?

Medos has received an increase in complaints for CEREBASE DA Guide Sheath with reports of fractures at the distal end, which may result in surgical procedural delay, vascular injury and in extreme rare occasions it may result in embolism.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

a) GS9080SD: UPDATED 4/4/2024 UDI-DI 10886704082316, Lot numbers: 31121042, 31133636, 31140739, 31140751, 31146719, 31154809, 31160736, 31174927, 31178955, 31189206, 31194710, 31194733, 31208988, 31212653, 31212661; b) GS9090SD: UPDATED 4/4/2024 UDI-DI 10886704082293, Lot numbers: 31094249, 31103843, 31108260, 31116484, 31116485, 31121039, 31121040, 31121043, 31121044, 31133637, 31133638, 31140740, 31140952, 31146720, 31150968, 31150969, 31154810, 31154811, 31160737, 31160738, 31168610, 31168611, 31174928, 31174930, 31178956, 31178957, 31178958, 31189207, 31189208, 31194711, 31194712, 31194734, 31194735, 31204436, 31208989, 31208990, 31208991, 31212654, 31212655, 31212663, 31225738; c) GS9095SD: UDI-DI 10886704082323, Lot Numbers: 31208992, 31212656

Product
CERENOVUS CEREBASE DA Guide Sheath, Part Numbers: a) GS9080SD; b) GS9090SD; c) GS9095SD; Vascular guide-catheter, single-use
Recalling firm
Medos International Sarl
Quantity
1343 units
Distribution
Worldwide distribution.
Recall date
March 20, 2024
Classification
Class I
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls