QMS Tacrolimus Immunoassay recall
QMS Tacrolimus Immunoassay REF 10015573 CALIBRATOR SET QMS TACROLIMUS CAL A 10015566-3 QMS TACROLIMUS CAL B 10015567-3 QMS TACROLIMUS CAL C 10015568-3…
Why was QMS Tacrolimus Immunoassay recalled?
Due to QMS Tacrolimus Calibrator recovering between 82 to 94% of their target values, resulting in over-quantitation in tacrolimus patient result values when used with the QMS Tacrolimus Immunoassay.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Model/Part Number: 10015573 UDI Codes: (01) 0884883 01259 2 (17) 220310 (10) 74175462 (01) 0884883 01259 2 (17) 220512 (10) 74267412 Lot Numbers: 74175462 74267412 Calibrators within the set: QMS TACROLIMUS CAL A 10015566-3 QMS TACROLIMUS CAL B 10015567-3 QMS TACROLIMUS CAL C 10015568-3 QMS TACROLIMUS CAL D 10015569-3 QMS TACROLIMUS CAL E 10015570-3 QMS TACROLIMUS CAL F 10015571-3
- Product
- QMS Tacrolimus Immunoassay REF 10015573 CALIBRATOR SET QMS TACROLIMUS CAL A 10015566-3 QMS TACROLIMUS CAL B 10015567-3 QMS TACROLIMUS CAL C 10015568-3…
- Recalling firm
- Microgenics Corporation
- Quantity
- 615 kits
- Distribution
- U.S.: AL, AZ, CA, DC, FL, IA, ID, IL, KY, LA, MA, NE, NJ, NM, OH, OK, PA, SC, TN, TX, and VA O.U.S.: Bangladesh, Brazil, Canada, Germany, India, Malaysia, Mexico, South Korea, and Thailand
- Recall date
- July 6, 2022
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)