FDAClass IIApril 6, 2016

Estradiol II Elecsys and cobas e analyzers recall

Estradiol II Elecsys and cobas e analyzers REF 03000079 190; M 6.5mL, R1 8 mL, R2 8 mL, IVD, For USA: CONTENT M Streptavidin-coated microparticles…

Why was Estradiol II Elecsys and cobas e analyzers recalled?

Due to the risk of a recently identified cross reactivity, the Elecsys Estradiol assay should not be used when monitoring patient being treated with fulvestrant. Customers should be informed that fulvestrant will increase the apparent concentration of estradiol in women being treated with this drug. If treatment with fulvestrant has been altered or discontinued as a result of falsely elevated e

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Part Number Elecsys Estradiol II : 03000079190; all lot numbers are affected.

Product
Estradiol II Elecsys and cobas e analyzers REF 03000079 190; M 6.5mL, R1 8 mL, R2 8 mL, IVD, For USA: CONTENT M Streptavidin-coated microparticles…
Recalling firm
Roche Diagnostics Operations, Inc.
Quantity
102,312
Distribution
US Nationwide Distribution including Puerto Rico.
Recall date
April 6, 2016
Classification
Class II
Status
Terminated
Agency
FDA (device)