FDAClass IIMarch 12, 2025

MIM software recall: Software Issue

MIM software; System, Image Processing, Radiological

Why was MIM software recalled?

In situations where two images with differing Fields of View (FOV) complete an image fusion, an incorrect, elevated Maximum Standardized Uptake Value (SUV) within MIM Software versions 7.2.0 through 7.2.6.could result.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Software versions: 1. MIM Anyware 1.1, UDI/DI Not Applicable 2. MIM 7.2.0, UDI/DI 00850009343192; 3. MIM 7.2.1, UDI/DI 00850009343192; 4. MIM 7.2.2, UDI/DI 00850009343192; 5. MIM 7.2.3, UDI/DI 00850009343192; 6. MIM 7.2.4, UDI/DI 00850009343192; 7. MIM 7.2.5, UDI/DI 00850009343192; 8. MIM 7.2.6, UDI/DI 00850009343192

Product
MIM software; System, Image Processing, Radiological
Recalling firm
MIM Software Inc
Quantity
352 units
Distribution
Worldwide distribution - United States Nationwide and the countries of Australia, Belgium, Bolivia, Canada, China, Croatia, Czech Republic, Denmark, France, Germany, Greece, Israel, Italy, Japan, Mexico, Netherlands, Poland, Russia, Spain, Sweden, United Kingdom, Finland, Hong Kong.
Recall date
March 12, 2025
Classification
Class II
Status
Ongoing
Agency
FDA (device)