Siemens ADVIA Centaur Systems Calibrator A Kits recall
Siemens ADVIA Centaur Systems Calibrator A Kits, Cat No. 04800646, SMN 10285903 (2 pack)Cat No. 04800735, SMN 10285904 (6 pack) Cat No. 04800840, SMN…
Why was Siemens ADVIA Centaur Systems Calibrator A Kits recalled?
Siemens Healthcare Diagnostics has confirmed a negative bias for ADVIA Centaur FT4 when used with Calibrator A kit lots ending in 90 on the ADVIA Centaur, ADVIA Centaur XP and ADVIA Centaur XPT Systems. In addition, Siemens Healthcare Diagnostics confirmed the potential for calibration failures due to above limit calibrator RLU %CVs when using Calibrator A kit lots ending in 90 with the FT4 assay. The performance of the FT4 assay when used with Calibrator A kit lots ending in 90 on the ADVIA Centaur CP System is not affected. Customers may continue to use Calibrator A kit lots ending in 90 for the ADVIA Centaur Systems FT3, T3, T4 and TUp assays.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Lot Numbers: 07995A90 08451A90 16035A90 32420A90 43871A90 59675A90 62739A90 Exp. Date: 2017/07/24
- Product
- Siemens ADVIA Centaur Systems Calibrator A Kits, Cat No. 04800646, SMN 10285903 (2 pack)Cat No. 04800735, SMN 10285904 (6 pack) Cat No. 04800840, SMN…
- Recalling firm
- Siemens Healthcare Diagnostics, Inc
- Quantity
- 9664 kits
- Distribution
- Nationwide. Afghanistan, Albania, Angola, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, French Polynesia, Georgia, Germany, Greece, Guadeloupe, Guatemala, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Latvia, Lithuania, Luxembourg, Macedonia, Malaysia, Malta, Martinique, Mexico, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Republic of Korea, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Turkmenistan, United Arab Emirates, United Kingdom, Uruguay, Vatican, Vietnam
- Recall date
- March 8, 2017
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)