Carolina(r) Liquid Chemistries Corp. Etha Rgt. KIT recall
CAROLINA(R) LIQUID CHEMISTRIES CORP. ETHA RGT. KIT, IVD, REF ST321
Why was Carolina(r) Liquid Chemistries Corp. Etha Rgt. KIT recalled?
Supplier testing showed that the absorbance rate of the Ethyl Alcohol Assay did not meet internal quality control specifications. The defective lots will not function toward the end of shelf life.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Lot 11687
- Product
- CAROLINA(R) LIQUID CHEMISTRIES CORP. ETHA RGT. KIT, IVD, REF ST321
- Recalling firm
- Carolina Liquid Chemistries
- Quantity
- 20 kits
- Distribution
- NY. NH, ME, NC, TX, AR, SC, CT, IL, CA, NJ
- Recall date
- May 8, 2019
- Classification
- Class III
- Status
- Terminated
- Agency
- FDA (device)