EliA Sample Diluent recall
EliA Sample Diluent, Article number, 83-1023-01
Why was EliA Sample Diluent recalled?
Erroneous EliA test results related to problems with liquid level detection and diluent pipetting in Phadia 250 instruments, and deformed bottles. Investigation showed problems with the liquid level detection in diluent bottle batch JJCA resulting in aspiration of no or too low volumes of liquid from the diluent bottle and from the dilution plate.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
EliA Sample Diluent, Article number, 83-1023-01, lots JM3Y, JS0B, JP87, JK5N, and JU9C
- Product
- EliA Sample Diluent, Article number, 83-1023-01
- Recalling firm
- Phadia US Inc
- Quantity
- 5957
- Distribution
- Domestic: AR, CA, CO, FL, GA, IL, IN, KY, LA, MI, MN, MO, MT, NC, NH, NJ, NM, NY, OH, OR, PA, SD, TN, TX, WI, WV Foreign: None VA/DOD: None
- Recall date
- March 8, 2017
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)