FDAClass IIMarch 22, 2023

TFN-Advanced Femoral Nail 12MM/130 DEG TI CANN TFNA recall

TFN-Advanced Femoral Nail 12MM/130 DEG TI CANN TFNA 170MM STERILE. Intended for temporary fixation and stabilization of proximal femur fractures Part…

Why was TFN-Advanced Femoral Nail 12MM/130 DEG TI CANN TFNA recalled?

TFNA Femoral Nails labeled and etched as 200mm may measure 170mm in length, and TFNA Femoral Nails labeled and etched as 170mm may measure 200mm in length

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI-DI: 10886982097262 Lot Number: 610P788

Product
TFN-Advanced Femoral Nail 12MM/130 DEG TI CANN TFNA 170MM STERILE. Intended for temporary fixation and stabilization of proximal femur fractures Part…
Recalling firm
Synthes (USA) Products LLC
Quantity
6 units
Distribution
Worldwide - US Nationwide distribution in the states of MS, OR, TX and the countries of Canada, Japan.
Recall date
March 22, 2023
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls