FDAClass IIFebruary 11, 2026

Volk Single Use Vitrectomy Lenses 1. Flat Vitrectomy recall: Sterility Concern

Volk Single Use Vitrectomy Lenses 1. Flat Vitrectomy Lens, Catalog VFD 2. Volk Magnifying Lens, Catalog VMD

Why was Volk Single Use Vitrectomy Lenses 1. Flat Vitrectomy recalled?

There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

1. Flat Vitrectomy Lens Pouch UDI 810123482290 Box UDI 810123482306 Lots 2410036 2411020 2. Volk Magnifying Lens Pouch UDI 810123482825 Box UDI 810123482832 Lot 2411019

Product
Volk Single Use Vitrectomy Lenses 1. Flat Vitrectomy Lens, Catalog VFD 2. Volk Magnifying Lens, Catalog VMD
Recalling firm
Vortex Surgical Inc.
Quantity
14,789 (8651 US; 6138 OUS)
Distribution
Worldwide Distribution: US (nationwide) to states of: AL, AR, AZ, CA, CT, FL, HI, IA, IL, IN, KY, MA, MI, MN, MS, NC, NE, NJ, NM, NY, OH, OR, PA, RI, SC, SD, TN, TX, WA, WI, VA; and OUS (foreign) to countries of: Australia, Chile, China, France, Germany, India, Ireland, Italy, Japan, Korea, Mexico, Netherlands, Philippines, Poland, Puerto Rico, Spain, Switzerland, Taiwan, United Kingdom, Vietnam.
Recall date
February 11, 2026
Classification
Class II
Status
Ongoing
Agency
FDA (device)