FDAClass IMarch 12, 2025

Rotarex Atherectomy System and Instructions for Use recall

Rotarex Atherectomy System and Instructions for Use, BD SET Rotarex: S 6 F x 110 cm, REF: 80236; S 6 F x 135 cm, REF: 80237; S 8 F x 85 cm, REF…

Why was Rotarex Atherectomy System and Instructions for Use recalled?

Atherectomy Catheter eIFU updated to clarify and emphasize procedural steps intended to reduce the likelihood of catheter breakage events. Catheter has an outer cylinder connected to a rotating helix, which could fracture and/or break, which would require retrieval, and helix facture/break could cause vessel injury and lead to severe bleeding.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Instructions for Use versions prior to ZE10895 revision C1 09/24. REF/UDI-DI/Lot(Expiration): 80236/07640142811855/230893(2025-12-15), 231100(2026-06-09), 231172(2026-06-09), 231171(2026-06-09), 231526(2026-08-21), 231875(2026-08-21), 240018(2026-11-22), 240017(2026-11-22), 240375(2027-01-18), 240567(2027-01-18), 241075(2024-05-22), 241335(2024-06-19), 241550(2024-07-19), 241777(2024-08-19), 242691(2024-12-13); 80237/07640142811862/230897(2026-04-11), 230898(2026-04-11), 231033(2026-05-31), 231101(2026-06-13), 231170(2026-06-13), 231229(2026-07-07), 231400(2026-07-14), 231414(2026-07-31), 231412(2026-06-30), 231413(2026-07-31), 231411(2026-08-03), 231521(2026-09-01), 231520(2026-07-31), 231522(2026-08-31), 231523(2026-09-01), 231524(2026-09-01), 231525(2026-09-01), 231715(2026-09-22), 231885(2026-09-22), 231884(2026-10-27), 240020(2026-11-22), 240396(2026-12-08), 240530(2027-01-10), 240531(2027-01-10), 240566(2027-01-10), 240568(2027-01-25), 240754(2027-02-06), 240769(2027-02-09), 240795(2027-03-04), 241074(2027-03-07), 241119(2027-03-21), 241190(2027-04-16), 241325(2027-04-16), 241328(2027-04-25), 241336(2027-04-29), 241464(2027-05-07), 241471(2027-05-27), 241553(2027-06-05), 242019(2027-06-28), 242023(2027-06-12), 242057(2027-06-20), 242056(2027-06-20), 242112(2027-06-26), 242160(2027-07-25), 242321(2027-07-08), 242493(2027-08-01), 242596(2027-09-20), 242693(2027-11-06); 80238/07640142811879/230895(2025-11-23), 240128(2026-05-31), 240276(2026-05-31), 240532(2026-05-31), 240658(2026-09-30), 241076(2026-09-30), 241330(2026-09-30), 241548(2026-09-30), 241768(2026-09-30), 241911(2026-09-30), 242117(2026-09-30), 242494(2026-09-30), 242521(2026-09-30), 242519(2026-05-31), 242520(2026-09-30); 80239/07640142811886/230896(2025-11-23), 231231(2026-06-13), 231410(2026-06-13), 231527(2026-08-31), 231883(2026-10-19), 232029(2026-08-31), 232028(2026-10-19), 240129(2026-11-30), 240278(2026-11-30), 240557(2027-01-22), 240746(2027-01-22), 240794(2027-02-28), 241077(2027-04-08), 241188(2027-04-08), 241462(2027-05-03), 241523(2027-05-03), 241767(2027-06-26), 241910(2027-06-26), 242159(2027-08-09), 242322(2027-09-10), 242518(2027-09-10), 242694(2027-10-25), 242695(2027-10-25)

Product
Rotarex Atherectomy System and Instructions for Use, BD SET Rotarex: S 6 F x 110 cm, REF: 80236; S 6 F x 135 cm, REF: 80237; S 8 F x 85 cm, REF…
Recalling firm
Bard Peripheral Vascular Inc
Quantity
15,755
Distribution
US Nationwide Distribution: NY, VA, KY, IN, OR, FL, MD, NJ, LA, CO, PR, PA, MS, MI, TX, KS, IL, CA, GA, OH, AL, NC, ND, AZ, SC, TN, UT, AR, WI, SD, MA, MO, ME, DE, WV , NV, NH, MN, CT, NE, IA, NM, WA, HI, and OK
Recall date
March 12, 2025
Classification
Class I
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls