Siemens ADVIA Centaur Enhance Estradiol Test Code eE2 recall
Siemens ADVIA Centaur Enhance Estradiol Test Code eE2 Catalog Number: 10490889, 10491445, 10697757 Product Usage: For in vitro diagnostic use in the…
Why was Siemens ADVIA Centaur Enhance Estradiol Test Code eE2 recalled?
The drug fulvestrant (Faslodex¿) may cause falsely elevated estradiol results.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
All lots
- Product
- Siemens ADVIA Centaur Enhance Estradiol Test Code eE2 Catalog Number: 10490889, 10491445, 10697757 Product Usage: For in vitro diagnostic use in the…
- Recalling firm
- Siemens Healthcare Diagnostics, Inc
- Quantity
- 33604 units
- Distribution
- Worldwide Distribution US Nationwide , AR, AU, AU, BD, BO, BR, CA, CH, CL, CN, CO, DO, EC, EG, HK, ID, IE, IL, IN, JP, KR, LK, LY, MX, MY, NI, NP, NZ, PE, PH, PY, SA, SG, SV, TH, TW, UY, VE, VN, and ZA.
- Recall date
- March 30, 2016
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)