FDAClass IIMarch 11, 2015

The I-Cart is used to store recall

The I-Cart is used to store, transport, and power video equipment used in medical procedures.

Why was The I-Cart is used to store recalled?

The transformers installed on carts used for Richard Wolf medical video systems may overheat causing smoke and black plastic material to leak from the cart chassis.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

(NOTE: The listed serial numbers are for the carts manufactured and include those that were distributed.) 1) Part Number: 31113.801; Serial Numbers: 1306000, 1306007, 1306014, 1306021, 1306103, 1306001, 1306008, 1306015, 1306022, 1306104, 1306002, 1306009, 1306016, 1306023, 1306105, 1306003, 1306010, 1306017, 1306024, 1307100, 1306004, 1306011, 1306018, 1306100, 1307101, 1306005, 1306012, 1306019, 1306101, 1307102, 1306006, 1306013, 1306020, 1306102, 1317103. 2) Part Number: 31113.701; Serial Numbers: 11001, 11011, 11020, 11030, 11039, 11048, 11058, 11002, 11012, 11021, 11031, 11040, 11049, 11059, 11003, 11013, 11023, 11032, 11041, 11050, 11060, 11005, 11014, 11024, 11033, 11042, 11051, 11061, 11006, 11015, 11025, 11034, 11043, 11052, 11063, 11007, 11016, 11026, 11035, 11044, 11053, 11064, 11008, 11017, 11027, 11036, 11045, 11055, 11065, 11009, 11018, 11028, 11037, 11046, 11056, 11066, 11010, 11019, 11029, 11038, 11047, 11057, 11067.

Product
The I-Cart is used to store, transport, and power video equipment used in medical procedures.
Recalling firm
Richard Wolf Medical Instruments Corp.
Quantity
83 carts
Distribution
Nationwide
Recall date
March 11, 2015
Classification
Class II
Status
Terminated
Agency
FDA (device)