FDAClass IIMarch 15, 2023

Sensus Healthcare Inc recall

Sensus Healthcare Inc, SRT-100 Vision IPX 0, HFUS Module IPX 1, with software version 1.9.2; Superficial X-ray Radiation Therapy System with…

Why was Sensus Healthcare Inc recalled?

When the user pauses the beam delivery, the dose timer is reset (to zero), and as a result may deliver an unexpected dose to the patient.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Software version 1.9.2 loaded onto Serial Numbers: 1512-2002; 1512-2003; 1606-2007; 1609-2010; 1609-2011; 1609-2018; 1612-2023; 1612-2025; 1705-2032; 1709-2046; 1803-2068; 1906-2137; 1909-2149; 1912-2167; 2003-2177; 2112-2201; 2112-2202; 2112-2208; 2112-2218; 2203-2232; 2203-2241; 2206-2264; 2206-2272

Product
Sensus Healthcare Inc, SRT-100 Vision IPX 0, HFUS Module IPX 1, with software version 1.9.2; Superficial X-ray Radiation Therapy System with…
Recalling firm
Sensus Healthcare, Inc.
Quantity
23 units
Distribution
US Distribution to states of: RX, IN, NM, FL, NY, FA, NC, IA, and IL.
Recall date
March 15, 2023
Classification
Class II
Status
Terminated
Agency
FDA (device)