FDAClass IIMarch 5, 2025

DxC 500 AU Clinical Chemistry Analyzer recall: Software Issue

DxC 500 AU Clinical Chemistry Analyzer, REF: C63519, and C63520 with affected software.

Why was DxC 500 AU Clinical Chemistry Analyzer recalled?

Beckman Coulter has identified an issue which prevents the DxC 500 AU instrument from providing the correct onboard stability (Open Expiration) dates for assays that were loaded prior to a software upgrade (e.g. 1.3 to 1.4, 1.4 to 1.4.1, etc.). As a result, assays that have expired may still be used for sample processing, leading to inaccurate or erroneous patient test results. The issue only happens on assays loaded before instruments doing upgrades and does not impact new installation. It does not impact the assays loaded after upgrade.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Software Versions: V1.4, V1.4.1, V1.4.2, V1.4.3 REF/UDI-DI/Serial Numbers: C63519/14987666545058/2024090170; C63520/14987666545065/2024070116, 2023070009, 2024010038, 2024010035, 2024050072, 2024040056, 2024030049, 2024020046, 2024070111, 2024030051, 2024030052, 2024010033, 2023090015, 2024060103, 2024060104, 2024060101, 2024080145, 2024020045, 2024020042, 2024020043, 2024020047, 2024060096, 2024040059, 2023100019, 2024030053, 2023070010, 2024070129, 2024070130, 2024020041, 2024090183, 2024010036, 2024010037, 2024070132, 2024070131, 2023100020, 2024050064, 2024050070, 2023090016, 2023110022, 2024070110, 2024060075, 2024060106, 2023110023, 2024010024, 2024010025, 2024050066, 2024010029, 2024060107, 2023080012, 2024040060, 2024050074, 2024040062, 2024010031, 2024030055, 2024080144

Product
DxC 500 AU Clinical Chemistry Analyzer, REF: C63519, and C63520 with affected software.
Recalling firm
BECKMAN COULTER LABORATORY SYSTEMS (SUZHOU) CO., LTD.
Quantity
56
Distribution
US: MA, CA, SC, LA, OK, GA, AL, TX, OR, ID, CO, MN, NY, MS, PA, WA, IL, MI, PR. OUS: Lebanon, Poland, Ghana, Spain, South Africa, Slovakia, Ghana, Croatia, Seychelles, Australia, Republic of Korea
Recall date
March 5, 2025
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls