FDAClass IIMarch 30, 2016

cobas p 512 pre-analytical system Pre-analytical sample recall

cobas p 512 pre-analytical system Pre-analytical sample handling that includes de-capping, aliquoting and sorting of samples for analysis.

Why was cobas p 512 pre-analytical system Pre-analytical sample recalled?

Due to a false triggering or detection of the lifting gripper READY signal, sample tubes are not correctly placed back in the Rack Tube Transport (RTT) after the decapping process. Therefore, open sample tubes can be dropped in the cobas p 512, spilling the sample material.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Part numbers:05083435001 and 06268854001

Product
cobas p 512 pre-analytical system Pre-analytical sample handling that includes de-capping, aliquoting and sorting of samples for analysis.
Recalling firm
Roche Diagnostics Operations, Inc.
Quantity
35 Units
Distribution
US Distribution including Puerto Rico and to the states of :TX, OH, TN, AZ, WA, MI and GA
Recall date
March 30, 2016
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls