FDAClass IJune 12, 2019

ECMO Reinforced Dual Lumen Cannula recall

ECMO Reinforced Dual Lumen Cannula, Sterile EO, Rx Only, REF: VV19F

Why was ECMO Reinforced Dual Lumen Cannula recalled?

Clamping the clear tube very close to the blue hub may cause a localized adhesive failure in the bond between the clear tube and the molded hub. Forcefully rotating the clamp while it is in place may potentially cause an immediate disconnect of the clear tube or lead to a later fatigue failure of the adhesive bond, with subsequent clear tube disconnect from the hub.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lots: N18447-1A, P18909, P18971

Product
ECMO Reinforced Dual Lumen Cannula, Sterile EO, Rx Only, REF: VV19F
Recalling firm
OriGen Biomedical, Inc.
Quantity
61
Distribution
US: NY, AB, AR, FL, MO, NC, GA, LA, CO, AL, MA, CA, MI, WI, TX, DC, OH, CT, TN, IN, OR, IL, MN, QC, SC, KY, OK, AZ, UT, RI, NJ, SD, PA, MS, NM. Foreign (OUS): UAE, Saudi Arabia, Colombia, UK, Netherlands, Italy, Czech Republic, Germany, India, Japan, Canada, Thailand, Sweden, Belgium, Spain, Poland, South Africa, Chile
Recall date
June 12, 2019
Classification
Class I
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls