FDAClass IIMarch 15, 2023

Cardiosave Rescue Intra-Aortic Balloon Pump recall

Cardiosave Rescue Intra-Aortic Balloon Pump (IABP), Model Numbers 0998-00-0800-75, 0998-00-0800-83, and 0998-00-0800-85

Why was Cardiosave Rescue Intra-Aortic Balloon Pump recalled?

There have been reported failures of the high pressure helium regulator, which may cause a helium leak in the Cardiosave Hospital Cart. In instance of helium regulator failure, a Pump Console s internal reservoir of helium will not be replenished when docked into an impacted Hospital Cart. This may result in an insufficient amount of helium within the internal reservoir.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

All Serial Numbers. Model Number: 0998-00-0800-75, UDI-DI: 10607567112312; Model Number: 0998-00-0800-83, UDI-DI: 10607567108407; Model Number: 0998-00-0800-85, UDI-DI: 10607567113449

Product
Cardiosave Rescue Intra-Aortic Balloon Pump (IABP), Model Numbers 0998-00-0800-75, 0998-00-0800-83, and 0998-00-0800-85
Recalling firm
Datascope Corp.
Quantity
8909 units
Distribution
US Nationwide. Global Distribution.
Recall date
March 15, 2023
Classification
Class II
Status
Ongoing
Agency
FDA (device)