FDAClass IIMarch 15, 2023

CardioMEMS HF System PA Sensor and Delivery System recall

CardioMEMS HF System PA Sensor and Delivery System, Model Number CM3100, The CardioMEMS HF System provides pulmonary artery (PA) hemodynamic data used…

Why was CardioMEMS HF System PA Sensor and Delivery System recalled?

Select CardioMEMS PA Sensors (Model CM2000) operate outside of the intended frequency range at higher elevations when readings are taken over approximately 2,000 ft. (610 meters) above sea level; operation outside of the intended radiofrequency has the potential to result in inaccurate readings or sensor signal acquisition difficulties.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI/DI 05415067034724: M220200004, M220200005, M220200006, M220200008, M220200011, M220200014, M220200016, M220200017, M220200018, M220200019, M220200021, M220200022, M220200026, M220200028, M220200029, M220200031, M220200032, M220200034, M220200035, M220200036, M220200037, M220200039, M220200040, M220200041, M220200042, M220300002, M220300003, M220300004, M220300005, M220300007, M220300008, M220300009, M220300010, M220300011, M220300012, M220300013, M220300014, M220300016, M220300017, M220300018, M220300019, M220300021, M220300022, M220300023, M220300024, M220300025, M220300026, M220300027, M220300028, M220300029, M220300030, M220300031, M220300032, M220300035, M220300036, M220300040, M220300041, M220300042, M220300043, M220300044, M220300045, M220300046, M220300047, M220300048, M220300049, M220300050, M220300051, M220300052, M220300053, M220300054, M220300055, M220300057, M220300058, M220300059, M220300060, M220300061, M220300062, M220300063, M220300064, M220300065, M220300066, M220300067, M220300069, M220300070, M220300071, M220300073, M220300074, M220300075, M220300077, M220300078, M220300080, M220300081, M220300082, M220300083, M220300084, M220300085, M220300087, M220300088, M220300089, M220300091, M220300092, M220300093, M220300096, M220300097, M220300101, M220300102, M220300103, M220300104, M220300105, M220300107, M220300108, M220300109, M220300111, M220300112, M220300114, M220300115, M220300116, M220300118, M220300119, M220300120, M220300121, M220300122, M220300123, M220300125, M220300126, M220300127, M220300129, M220300131, M220300136, M220300138, M220300139, M220300140, M220300141, M220300142, M220300143, M220300144, M220300145, M220300146, M220300147, M220300150, M220300151, M220300152, M220300153, M220300154, M220300155, M220300156, M220300157, M220300158, M220300160, M220300161, M220300162, M220300163, M220300164, M220300166, M220300167, M220300168, M220300169, M220300170, M220300171, M220300172, M220300173, M220300174, M220300175, M220300176, M220300177, M220300178, M220300179, M220300180, M220300181, M220300182, M220300183, M220300184, M220300186, M220300187, M220300188, M220300189, M220300190, M220300191, M220300192, M220300193, M220300194, M220300195, M220300196, M220700025, M220700026, M220700027, M220700028, M220700031, M220700033, M220700034, M220700035, M220700036, M220700037, M220700041, M220700042, M220700043, M220700045, M220700051, M220700052, M220700054, M220700059, M220700061, M220700097, M220700098, M220700102, M220700104, M220700106, M220700107, M220700109, M220700110, M220700111, M220700112, M220700113, M220700114, M220700115, M220700116, M220700117, M220700118, M220700119, M220700120, M220700121, M220700122, M220700123, M220700124, M220700125, M220700126, M220700128, M220700129, M220700130, M220700131, M220700132, M220700133, M220700134, M220700135, M220700136, M220700137, M220700138, M220700139, M220700142, M220700143, M220700144, M220700145, M220700146, M220700147, M220900017, M220900018, M220900021, M220900023, M220900024, M220900025, M220900026, M220900030, M220900031, M220900032, M220900033, M220900034, M220900035, M220900037, M220900038, M220900040, M220900042, M220900044, M220900045, M220900046, M220900047, M220900048, M220900049, M220900050, M220900052, M220900063, M220900065, M220900066, M220900067, M220900068, M220900069, M220900070, M220900071, M220900072, M220900073, M220900075, M220900076, M220900078, M220900079, M220900080, M220900081, M220900086, M220900087, M220900088, M220900089, M220900090, M220900091, M220900093, M220900094, M220900095, M220900096, M220900097, M220900101, M220900104, M220900106, M220900109, M221000023, M221000030, M221000032, M221000034, M221000035, M221000038, M221000039, M221000041, M221000042, M221000044, M221000047, M221000048, M221000049, M221000050, M221000052, M221100002, M221000046

Product
CardioMEMS HF System PA Sensor and Delivery System, Model Number CM3100, The CardioMEMS HF System provides pulmonary artery (PA) hemodynamic data used…
Recalling firm
St. Jude Medical
Quantity
317 units
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Canada, Colombia, Czech Republic, Denmark, Finland, France, Germany, Ireland, Italy, Kuwait, Netherlands, Portugal, Saudi Arabia, Spain, Sweden, Switzerland, UK, United Arab Emirates.
Recall date
March 15, 2023
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls