FDAClass IIIFebruary 22, 2017

The Enzyme II Calibrator is an in vitro diagnostic recall

The Enzyme II Calibrator is an in vitro diagnostic product for the Alanine Aminotransferase (ALTI) on the Dimension¿ clinical chemistry system. For…

Why was The Enzyme II Calibrator is an in vitro diagnostic recalled?

Complaints of positive shift of patient and QC results when the ALTI assay is calibrated with the Dimension Enzyme II calibrator. Internal investigation has confirmed that patient samples and QC material demonstrate a shift of up to 10% in comparison to results obtained using earlier calibrator lots.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Product Lot: 4MD047, 5CD052, 5FD017, 5GD050, 5HD010, 5JD069, 5LD051, 6AD054

Product
The Enzyme II Calibrator is an in vitro diagnostic product for the Alanine Aminotransferase (ALTI) on the Dimension¿ clinical chemistry system. For…
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
24,940.00
Distribution
Distributed throughout the United States
Recall date
February 22, 2017
Classification
Class III
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls