FDAClass IMarch 26, 2014

FreeStyle Blood Glucose Monitors System. For in vitro recall

FreeStyle Blood Glucose Monitors System. For in vitro diagnostic use only. Store at room temperature below 86 degrees F (30 degrees C). Use test strips only within the system operating temperature range as outlined in your Owner's Booklet. Meter made in China; Distributed by Abbott Diabetes Care, Alameda, CA The FreeStyle mete should only be used with FreeStyle Test Strips and FreeStyle Control Solution. Using other brands of test strips and control solutions with the FreeStyle Mete can product inaccurate results. The FreeStyle Glucose meter is incorporated into the OmniPod insulin Management system (manufactured by Insulet Corporation, Bedford, MA and is intended for subcutaneous delivery of insulin. Intended to monitor blood glucose from samples taken from the body. IVD use only,

Why was FreeStyle Blood Glucose Monitors System. For in vitro recalled?

Abbott Diabetes Care has identified through internal testing and investigation that all non-applied voltage legacy meters have the potential to produce out of range control solution results and/or erroneously low blood glucose results when used in conjunction with Free Style test strips lots within expiry.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Part number CAT11001-0; all serial numbers

Product
FreeStyle Blood Glucose Monitors System. For in vitro diagnostic use only. Store at room temperature below 86 degrees F (30 degrees C). Use test strips only within the system operating temperature range as outlined in your Owner's Booklet. Meter made in China; Distributed by Abbott Diabetes Care, Alameda, CA The FreeStyle mete should only be used with FreeStyle Test Strips and FreeStyle Control Solution. Using other brands of test strips and control solutions with the FreeStyle Mete can product inaccurate results. The FreeStyle Glucose meter is incorporated into the OmniPod insulin Management system (manufactured by Insulet Corporation, Bedford, MA and is intended for subcutaneous delivery of insulin. Intended to monitor blood glucose from samples taken from the body. IVD use only,
Recalling firm
Abbott Diabetes Care, Inc.
Quantity
not provided. Device has not been shipped in the last 5 years.
Distribution
Worldwide Distribution: US (nationwide) and countries of: Austria, Canada, Germany, Israel, Netherlands, Norway, Sweden, Switzerland, United Kingdom.
Recall date
March 26, 2014
Classification
Class I
Status
Terminated
Agency
FDA (device)