Stryker Howmedica Osteonics Corp. AxSOS System recall
Stryker Howmedica Osteonics Corp. AxSOS System Instrument. The AxSOS Locking Plate System is designed to treat periarticular or intraarticular…
Why was Stryker Howmedica Osteonics Corp. AxSOS System recalled?
Stryker Orthopaedics has received a report indicating that the scale of the AxSOS Calibrated Drill Bit 4.3,,X263mm, flat , manufactured by Stryker Osteosynthesis, is incorrect. The incorrect scale could result in the surgeon selecting a screw that is 10mm longer than intended.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Catalog number 703615
- Product
- Stryker Howmedica Osteonics Corp. AxSOS System Instrument. The AxSOS Locking Plate System is designed to treat periarticular or intraarticular…
- Recalling firm
- Stryker Howmedica Osteonics Corp.
- Quantity
- 107 units
- Distribution
- Nationwide Distribution including the states of AL, AZ, IL, KY, LA, NJ, NC, FL, CA, NM, MO, TN, VA, CO, MA, TX, IN and MT.
- Recall date
- May 8, 2013
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)