FDAClass IIMarch 8, 2023

DRX-Compass/DR-FIT X-ray Systems with Firmware version recall

DRX-Compass/DR-FIT X-ray Systems with Firmware version 2.3.2.0/2.3.2.4. For use in obtaining diagnostic quality radiographic images to aid the…

Why was DRX-Compass/DR-FIT X-ray Systems with Firmware version recalled?

After pressing and releasing the Z-Axis Motorized buttons on the Tube Head Display, the Overhead Tube Crane (OTC) could potentially continue the movement unexpectedly to a certain distance which could potentially result in a injury

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI: 60889978621613 (Rayco) 60889978624188 (1049) US Serial Numbers: SP0763 50066 50082 50089 50135 50145 50143 50140D 50144 OUS serial numbers: SP0819 SP0687G2 SP0726 SP0746 SP0834 SP0837 S0001 S0170 S0698G2 1085 FS0564 S0578 S0744 SP0903 SP0600 SP1035 SP0930 SP1007 SP1037 SP0772 S0934G2 S0639G2 S1020G2 S0118 S0135 S0134 S0165 S0167 S0168 S0181 S0800G2 S0784G2 S0811G2 S0832G2 S0826G2 S0822G2 S1038G2 S0678G2 S0619G2 S0636G2 S0641G2 S0643G2 S0701G2 S0738G2 S0758G2 S0760G2 S0673G2 S0785G2 S0842G2 S0860G2 S0863G2 S0888G2 S0973G2 S1105G2 S1048G2 S1073G2 S1004G2 S1110G2 S0791G2 S0941G2 SP0900 S0625

Product
DRX-Compass/DR-FIT X-ray Systems with Firmware version 2.3.2.0/2.3.2.4. For use in obtaining diagnostic quality radiographic images to aid the…
Recalling firm
Carestream Health, Inc.
Quantity
72 units
Distribution
Worldwide distribution - US Nationwide distribution in the states of California, Idaho, Illinois, Massachusetts, Minnesota, Nevada, North Carolina, Texas Wisconsin and the countries of Argentina, Australia, Brazil, China, Finland, France, INDIA, ITALY, Korea, Malaysia, Poland, Republic of Korea.
Recall date
March 8, 2023
Classification
Class II
Status
Ongoing
Agency
FDA (device)