Siemens Digital Linear Accelerators of type: ARTISTE MV recall
Siemens Digital Linear Accelerators of type: ARTISTE MV System MEVATRON M2/Primus Mid-Energy PRIMUS PRIMUS HI ONCOR Impression ONCOR Impression Plus…
Why was Siemens Digital Linear Accelerators of type: ARTISTE MV recalled?
Advisory issued not to disable or bypass the AMP function. Disabling or bypassing the AMP function for reasons of comfort is not advisable may lead to patient or user injury due to collision with LINAC equipment disabling the Auto Field Sequence (AFS) Automatic Motion Protection (AMP) function implemented at the control of the Digital Linear Accelerator.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
SIEMENS Digital Linear Accelerators of type ARTISTE", ONCOR" and PRIMUS" running: - Control Console from software version 13.0.302 and higher - Control Console from software version 9.2.400 and higher - Controle Console from software version 11.0.400 and higher
- Product
- Siemens Digital Linear Accelerators of type: ARTISTE MV System MEVATRON M2/Primus Mid-Energy PRIMUS PRIMUS HI ONCOR Impression ONCOR Impression Plus…
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Quantity
- 120 US units
- Distribution
- Nationwide
- Recall date
- May 1, 2019
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)