FDAClass IIMarch 6, 2024

Ref No: INSHITH / Integra¿ Cranial Access Kit. For recall: Sterility Concern

Ref No: INSHITH / Integra¿ Cranial Access Kit. For access to the subarachnoid space or the lateral ventricles of the brain.

Why was Ref No: INSHITH / Integra¿ Cranial Access Kit. For recalled?

A defect in the sterile packaging (header bag) of the Cranial Access Kit failing the required packaging integrity testing criteria. Therefore, product sterility maybe compromised.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI: 10381780263968 Lot Number Exp: Date: 6631603 1-Sep-23 6631604 1-Sep-23 6710212 1-Sep-23 6710213 1-Sep-23 6788598 1-Sep-23 6788599 1-Dec-23 6788600 1-Jan-24 6896383 1-Dec-23 6896384 1-Jan-24 6896385 1-Jan-24 6955153 17-Mar-24 6955154 17-Mar-24 6955155 17-Mar-24 7000260 17-Mar-24 7000261 17-Mar-24 7000262 17-Mar-24 7028225 17-Mar-24 7028226 17-Mar-24 7028227 1-Apr-24 7073035 1-Apr-24 7073036 1-Apr-24 7073037 1-Apr-24 7124699 1-Apr-24 7124700 1-Apr-24 7124702 1-Apr-24 7249008 1-Jun-24 7249009 1-Jun-24 7257038 1-Jun-24 7257039 1-Jun-24 7257040 1-Jun-24

Product
Ref No: INSHITH / Integra¿ Cranial Access Kit. For access to the subarachnoid space or the lateral ventricles of the brain.
Recalling firm
Integra LifeSciences Corp.
Quantity
8906 units
Distribution
Worldwide distribution - US Nationwide and the countries of Austria, Bahamas, Belgium, Canada, Chile, France, Germany, Greece, Ireland, Israel, Poland, Portugal, Slovenia, South Africa, Spain, United Arab Emirates (the), United Kingdom of Great Britain and Northern Ireland (the).
Recall date
March 6, 2024
Classification
Class II
Status
Ongoing
Agency
FDA (device)