stryker Hoffmann LRF recall
stryker Hoffmann LRF, Transport Strut, Model Number 4933-0-400; Orthopedic external fixation device intended to keep bone fragments/segments in a…
Why was stryker Hoffmann LRF recalled?
Stryker has identified a nonconformance in specific lots of Hoffmann LRF Bone Transport Struts. Specifically, the thread pitch is oversized at the distal end of the threaded rod, and as a result the distal ring fixation cannot be locked between 0- 50mm of the threaded rod.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Serial numbers: D32158, D32160, D32161, D32162, D32163, D32164, D34343, D34344, D40690, D41629, D41630, D41631, D44713, G16331, G16332, G16333, G26278, G29759, G29760, G32730, G32731, G32735, G45119, H30398, H30399, H66392
- Product
- stryker Hoffmann LRF, Transport Strut, Model Number 4933-0-400; Orthopedic external fixation device intended to keep bone fragments/segments in a…
- Recalling firm
- Stryker GmbH
- Quantity
- 488 units
- Distribution
- US Nationwide distribution.
- Recall date
- March 8, 2023
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)