FDAClass IIFebruary 22, 2017

LIMA Modular Revision Hip Stem recall

LIMA Modular Revision Hip Stem Model 428-01-050_110 Product Usage: The Modular Revision Femoral hip stem is made up of a modular stem coupled with a…

Why was LIMA Modular Revision Hip Stem recalled?

Lima Proximal Bodies were inadvertently re-sterilized. The safety screw which affixes the distal and proximal bodies of the stem includes a thread-locking plug made from UHMWPE, which is not approved for repeated gamma sterilization.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

1005528A, 1006882A, 1107305A, 1101798A, 1100565A, 1005531A, 1211716A, 1007408A, 1004038B, 1005533D, 1106805A, 1203471A, 1005533C, 1100565A, 1101795A, 1007079A, 1005532A, 1100560A

Product
LIMA Modular Revision Hip Stem Model 428-01-050_110 Product Usage: The Modular Revision Femoral hip stem is made up of a modular stem coupled with a…
Recalling firm
Encore Medical, Lp
Quantity
13 devices
Distribution
US Nationwide - US Nationwide in the states of MS, OH, RI, NY, OK, TX, CA, HI
Recall date
February 22, 2017
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls