FDAClass IIMarch 17, 2021

Power Injectable CT Port insertion kits - Product recall

Power Injectable CT Port insertion kits - Product Usage: is an implantable access device designed to provide repeated access to the vascular system.

Why was Power Injectable CT Port insertion kits - Product recalled?

The 5F Dignity CT Port kits were packaged with the incorrect size port. The kits include a 6.6F port instead of the correct 5F port. The inability to connect the catheter lumen to the port stem because of the incorrect size may not be noted until well into the procedure, after the catheter lumen is placed in the vessel. An alternate port will need to be obtained prolonging the procedure. A prolonged insertion procedure increases any risk associated with a surgical procedure.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Catalog Number: MRDP50AXS, Lot Number: MPDX180, UDI Number: 884908149753 (updated 3/11/2021 added lot number: MPCY660)

Product
Power Injectable CT Port insertion kits - Product Usage: is an implantable access device designed to provide repeated access to the vascular system.
Recalling firm
Medical Components, Inc dba MedComp
Quantity
96 pieces total (updated 3/11/2021 additional 72 pieces)
Distribution
US Nationwide distribution in the states of CA, DC, GA, IL, IN, MA, MI, NJ, NM, NY, OH, PA, TX. Updated 3/11/2021
Recall date
March 17, 2021
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls