Medi-Trace Cadence Adult Radiotransparent recall
Medi-Trace Cadence Adult Radiotransparent Defibrillation Electrode Product ID: 22550R. Intended for use in cardioversion, pacing, ECG monitoring, and…
Why was Medi-Trace Cadence Adult Radiotransparent recalled?
Defibrillation Electrodes may arc/spark result in thermal damage to the leadwire, which could render the device incapable of delivering the appropriate energy or shock to the patient
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Lot Numbers: 226543X, 230054X , 230814X, 232182X, 232805X, 234245X, 304667X, 301844X, 300446X, 305320X
- Product
- Medi-Trace Cadence Adult Radiotransparent Defibrillation Electrode Product ID: 22550R. Intended for use in cardioversion, pacing, ECG monitoring, and…
- Recalling firm
- Covidien LLC
- Quantity
- 133470 (sets of 2)
- Distribution
- Worldwide Distribution-USA (nationwide) and the countries of Korea, Japan, Canada, Australia, Brazil, Belgium, Singapore, and South Africa.
- Recall date
- May 8, 2013
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)