Cannabinoids DAU Calibrator Control Level 1 recall
Cannabinoids DAU Calibrator Control Level 1 (37.5 ng/mL) This solution is prepared from non-sterile urine and should be handled as potentially…
Why was Cannabinoids DAU Calibrator Control Level 1 recalled?
Product catalog #s 0007c Cannabinoid 37.5 ng/mL Control Level 1 and 0008c Cannabinoid 62.5 Control Level 2 of the lots mentioned may give decreased concentration readings towards the end of their shelf life due to use of dated but unexpired secondary stock used to create the products.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Catalog number 0007Cm; lot number 1211158R; Expiration May 14, 2014.
- Product
- Cannabinoids DAU Calibrator Control Level 1 (37.5 ng/mL) This solution is prepared from non-sterile urine and should be handled as potentially…
- Recalling firm
- Lin-Zhi International Inc
- Quantity
- 130 items
- Distribution
- US Distribution including the states of SC, NC, LA, FL, TX,, AL and GA and Internationally to Germany.
- Recall date
- March 19, 2014
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)