FDAClass IIMay 1, 2013

The Alere Cholestech LDX¿ high sensitivity C-Reactive recall

The Alere Cholestech LDX¿ high sensitivity C-Reactive Protein (hsCRP) Test Cassette, Model Number 12-807. Product Usage: Immunoassay for the…

Why was The Alere Cholestech LDX¿ high sensitivity C-Reactive recalled?

The Alere Cholestech LDX¿ hsCRP cassette (PN 12-807) may have increased imprecision relative to the performance data in the package insert, which may cause differences in quantitative results for hsCRP compared to a reference method or repeat testing on the same patient sample.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Model Number 12-807. ¿ Lot/Unit Codes: 274632, 281847 ¿ Expiration date, or Expected shelf life: 5/5/13, 6/30/13

Product
The Alere Cholestech LDX¿ high sensitivity C-Reactive Protein (hsCRP) Test Cassette, Model Number 12-807. Product Usage: Immunoassay for the…
Recalling firm
Alere San Diego, Inc.
Quantity
2333 kits
Distribution
USA Nationwide Distribution
Recall date
May 1, 2013
Classification
Class II
Status
Terminated
Agency
FDA (device)