FDAClass IIMarch 19, 2014

PENTARAY NAV High-Density Mapping Catheter and PENTARAY recall

PENTARAY NAV High-Density Mapping Catheter and PENTARAY NAV ECO Catheter, Catalog No. D128201, D128202, D128204, D128205, D128207, D128208, D128210…

Why was PENTARAY NAV High-Density Mapping Catheter and PENTARAY recalled?

Biosense Webster is recalling the PentaRay Nav Catheter because it has the potential to have partial tip separation from the catheter shaft. Biosense Webster is expanding the recall on 5/1/14 to include all lots of the PentaRay Nav Catheter and the PentaRay Nav ECO Catheter because they can lead to an insufficient bond of the distal tip to the catheter shaft.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot Numbers: All Lots

Product
PENTARAY NAV High-Density Mapping Catheter and PENTARAY NAV ECO Catheter, Catalog No. D128201, D128202, D128204, D128205, D128207, D128208, D128210…
Recalling firm
Biosense Webster, Inc.
Quantity
15,672 units total (12,804 units in US)
Distribution
US Nationwide Distribution.
Recall date
March 19, 2014
Classification
Class II
Status
Terminated
Agency
FDA (device)