FDAClass IIMarch 19, 2014

Device Name: ABX PENTRA ML recall: Software Issue

Device Name: ABX PENTRA ML Model/Part #(s): A11A01652 Lot/Serial #(s): N/A Software Version(s): All versions are subject to this correction. Product…

Why was Device Name: ABX PENTRA ML recalled?

HORIBA Medical initiated this recall of all software versions of ABX PENTRA ML (Model A11A06152) due to a software bug with the potential of transmitting incorrect results to the Laboratory Information System.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Model/Part #(s): A11A01652 Lot/Serial #(s): N/A Software Version(s): All versions are subject to this correction. Shelf Life: N/A

Product
Device Name: ABX PENTRA ML Model/Part #(s): A11A01652 Lot/Serial #(s): N/A Software Version(s): All versions are subject to this correction. Product…
Recalling firm
Horiba Instruments, Inc dba Horiba Medical
Quantity
15
Distribution
USA Nationwide Distribution
Recall date
March 19, 2014
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls