VerifyNow IIb/IIIa Test recall
VerifyNow IIb/IIIa Test, Catalog No. 85310, 10-Test Kit and Catalog No. 85011, 25-Test kit. Product Usage: The VerifyNow System is a turbidimetric…
Why was VerifyNow IIb/IIIa Test recalled?
Accumetrics is recalling the VerifyNow IIb/IIIa test because it may result in the reporting of an erroneous low platelet aggregation unit (PAU) result. An erroneous low PAU result may cause a low percent inhibition calculation or low baseline PAU.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Lot/Serial No. WC0182B, WC0182C, WC0182A, WC0182D, WC0182E, WC0182F, WC0182G, WC0180C, WC0180E, WC0180F, WC0181A, WC0181C, WC0181D, WC0181F, WC0180A, WC0180B, WC0180D, WC0181B, WC0181E, WC0181G, WC0181H, WC0181J, WC0181K.
- Product
- VerifyNow IIb/IIIa Test, Catalog No. 85310, 10-Test Kit and Catalog No. 85011, 25-Test kit. Product Usage: The VerifyNow System is a turbidimetric…
- Recalling firm
- Accumetrics Inc
- Quantity
- 576 units total (524 units in US)
- Distribution
- USA Nationwide Distribution
- Recall date
- March 19, 2014
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)