FDAClass IMarch 10, 2021

ACIST Kodama Intravascular Ultrasound Catheter The recall

ACIST Kodama Intravascular Ultrasound Catheter The Kodama Intravascular Ultrasound Catheter is a component of the ACIST HDi System. The ACIST HDi…

Why was ACIST Kodama Intravascular Ultrasound Catheter The recalled?

Test results from the manufacturing line found a piece of damaged o-ring in an unexpected section of the catheter. Further testing indicated that pieces (>200 micron) of damaged o-ring had the potential to be flushed out of the catheter. ACIST is confirming the source of the failure mode to assure the quality and reliability of the Kodama catheter. There have been no related field reports related to this incident, nor any evidence or report of patient injury or adverse health consequence.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Model Number: 017788, 018125 (Japan only); Lot codes: 00233370 (100 units), 00233371 (90 units), 00233372 (100 units), 00233373 (100 units), 00233374 (100 units), 00233380 (100 units), 00233384 (60 units), 00233385 (100 units), 00233393 (100 units), 00233394 (100 units), 00237604 (35 units), 00237613 (100 units), 03012517 (100 units)

Product
ACIST Kodama Intravascular Ultrasound Catheter The Kodama Intravascular Ultrasound Catheter is a component of the ACIST HDi System. The ACIST HDi…
Recalling firm
Acist Medical Systems
Quantity
1185 units
Distribution
Worldwide - US Nationwide Distribution in the states of AL, AR, CA, CO, FL, IL, KS, KY, LA, MD, MI, MO, NC, NJ, NY, OK, PA, RI, TN; In the countries of India, Italy, Japan, Poland, and United Arab Emirates.
Recall date
March 10, 2021
Classification
Class I
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls