Hamilton Heater Shaker recall
HAMILTON HEATER SHAKER, 115 VAC/ 230 VAC, 750 VA, Power Consumption 41V/ 100W (max.), Relative Humidity 15% - 85% no condensation. REF 19903400
Why was Hamilton Heater Shaker recalled?
There is a risk that the device can start a fire if flammable liquids contact and enter the device.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Various Part Numbers: Hamilton Company PART NUMBER/ DESCRIPTION 188318/ HHS2,2MM ORBIT,CUSTOMIZED; 188319/ HHS2,3MM ORBIT,CUSTOMIZED; 199000/ HHS 0.5 MTP FLAT BOTTOM; 199001/ HHS 2.0 MTP FLAT BOTTOM; 199002/ PLEASE USE PN: 199034; 199007/ HHS 2.0 NUNC DWP 96 2ML; 199008/ HHS 3.0 NUNC DWP 96 2ML; 199009/ HHS 1.5 MTP FLAT BOTTOM; 199011/ HHS,1.5MM,NUNC,96DWP,2ML; 199027/ HHS 1.5 MM SARSTEDT 48X1.5ML; 199033/ HHS2,2MM ORBIT,MTP FLT BOTTOM; 199034/ HHS2,3MM ORBIT,MTP FLT BOTTOM; 199037/ HHS 1.5MM NUNC DWP 96 2ML; 199038/ HHS2,2MM ORBIT,NUNC DWP96 2M; 199039/ HHS2,3MM ORBIT,NUNC DWP96 2M; 199085/ HHS 3.0 FLIPTUBE RACK; 199474/ HHS 3.0 BIO-RAD DWP FSD; 809018/ HHS 3.0 AUTOLYS; 0201710-001/ HHSADVWR40002-012DWP96CON2ML; 0202697-002/ HHS,RGT TRUGH HOLD FRAME,LID; 0203408-002/ HHS,199034; 188219APE/ HHS 3.0MM AMPLITUDE; 199033R/ HHS2,2MM ORBIT,MTP FLT BOT REF; 199034R/ HHS2,3MM ORBIT,MTP FLT BTM; 65608-01/ HHS2,3MM ORBIT, MPT; 65608-01R/ HHS2,3MM ORBIT, MPT,REFURB; 6601885-01/ KIT,HHS,3 MM ORBIT,ML PREP; 6601885-02/ KIT,HHS,2 MM ORBIT,ML PREP; 95379-01/ HHS2,3MM ORBIT,DWP96,EXACT SCI; Hamilton Bonaduz AG PART NUMBER DESCRIPTION 188319/ Hamilton Heater Shaker, flat bottom adapter (APE); 199033/ Hamilton Heater Shaker, flat bottom adapter; 199034/ Hamilton Heater Shaker, flat bottom adapter; 199037/ HHS 1.5 NUNC DWP 96 2ML; 199038/ HHS 2.0 NUNC DWP 96 2ML; 199039/ Hamilton Heater Shaker 3.0 Nunc DWP; 199068/ HHS 1.5 ROCHE; 199474/ HHS 3.0mm BIO-RAD FSD; 199525/ HHS 2.0 MTP V BOTTOM; 809018/ HHS 3.0 AUTOLYS; 816040/ HHS 1.5mm 2G; 816061/ HHS 1.5 NUNC DWP 96 2ML TIG; 819273/ HHS 3.0 V-BOTTOM; 10069018/ HHS 3.0 NUNC DWP 96 2ML TIG
- Product
- HAMILTON HEATER SHAKER, 115 VAC/ 230 VAC, 750 VA, Power Consumption 41V/ 100W (max.), Relative Humidity 15% - 85% no condensation. REF 19903400
- Recalling firm
- Hamilton Co
- Quantity
- 4729 units
- Distribution
- US - AL, AR, AZ, CA, CO, CT, DC. DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA. MD, ME. MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI AND WY OUS - AUSTRALIA, AUSTRIA, BRAZIL , CANADA , CHILE, CHINA, , COLOMBIA , FINLAND , GERMANY , ITALY , JAPAN , JPN , KOREA, SOUTH, PARAGUAY , PERU , PHL , ROMANIA , ROU , SINGAPORE, SWITZERLAND , SWITZERLAND , THAILAND, TWN , UNITED KINGDOM, AND URUGUAY
- Recall date
- March 10, 2021
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)