FDAClass IIMarch 19, 2014

Philips HeartStart MRx Monitor/Defibrillator recall

Philips HeartStart MRx Monitor/Defibrillator Model numbers: M3535A, M3536A, M3536M,M3536M4,M3536M6,861288,861289,861464,861483,and 861491

Why was Philips HeartStart MRx Monitor/Defibrillator recalled?

MRx may display a Red X in the Ready For Use (RFU) indicator during automated tests informing the user that a failure has been detected that may prevent the delivery of a shock and pacing

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Serial numbers within the range: US00100204 to US00570921

Product
Philips HeartStart MRx Monitor/Defibrillator Model numbers: M3535A, M3536A, M3536M,M3536M4,M3536M6,861288,861289,861464,861483,and 861491
Recalling firm
Philips Medical Systems, Inc.
Quantity
3541 units
Distribution
Nationwide Foreign: AUSTRIA BAHRAIN BELGIUM BRAZIL CANADA CHINA EGYPT ESTONIA FINLAND FRANCE GERMANY GHANA ICELAND INDIA INDONESIA IRELAND ISRAEL ITALY JAPAN JORDAN KENYA KOREA, REPUBLIC OF LATVIA LEBANON LIBYAN ARAB JAMAHIRIYA LITHUANIA MALAYSIA MEXICO MOZAMBIQUE MYANMAR (Burma) NETHERLANDS NETHERLANDS ANTILLES NEW CALEDONIA NEW ZEALAND NORWAY PAKISTAN PERU POLAND PORTUGAL QATAR RUSSIAN FEDERATION SAUDI ARABIA SINGAPORE SOUTH AFRICA SPAIN SWEDEN SWITZERLAND TAIWAN, PROVINCE OF CHINA THAILAND TURKMENISTAN UNITED ARAB EMIRATES UNITED KINGDOM VIET NAM
Recall date
March 19, 2014
Classification
Class II
Status
Terminated
Agency
FDA (device)