FDAClass IIFebruary 28, 2024

EndoVive Replacement Button Kit w/ ENFIT 28F 2.8 CM recall

EndoVive Replacement Button Kit w/ ENFIT 28F 2.8 CM, Outer Box UPN M00509500 or M00509510.

Why was EndoVive Replacement Button Kit w/ ENFIT 28F 2.8 CM recalled?

Affected batches of the kit contain Stoma Measuring Devices sized to measure for an 18/24F replacement button, but are labeled as being for a 28F. Use of defective product may lead to undersizing the button, which may cause embedding with erosion into the gastric wall, tissue necrosis, infection, sepsis, and associate sequelae.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UPN M00509500, GTIN (UDI-DI) 08714729902713, Lot 31073152; UPN M00509510, GTIN (UDI-DI) 08714729902720, Lots 31221751, 31761882, 32409573.

Product
EndoVive Replacement Button Kit w/ ENFIT 28F 2.8 CM, Outer Box UPN M00509500 or M00509510.
Recalling firm
Boston Scientific Corporation
Quantity
0 US; 74 OUS
Distribution
International distribution to the country of Japan only.
Recall date
February 28, 2024
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls