FDAClass IIMarch 28, 2018

Artis one fluoroscopic x-ray system and patient table recall

Artis one fluoroscopic x-ray system and patient table Artis one is an angiography system developed for diagnostic imaging and interventional…

Why was Artis one fluoroscopic x-ray system and patient table recalled?

Certain patient table systems may have a missing or incorrectly installed snap ring that could cause the bolt to slide out of its base and the tabletop carriage to slide to the floor.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Serial numbers: 82044 82105 82070 82213 82054 82040 82038 82205 82065 82267 82108 82010 82039 82109 82102 82066

Product
Artis one fluoroscopic x-ray system and patient table Artis one is an angiography system developed for diagnostic imaging and interventional…
Recalling firm
Siemens Medical Solutions USA, Inc
Quantity
16
Distribution
US Distribution to the states of : AL, AR, CA, FL, IN, LA, MI, MO, NC, NY, PA, TX, and WA.
Recall date
March 28, 2018
Classification
Class II
Status
Terminated