FDAClass IIFebruary 21, 2024

HeartSine samaritan PAD recall

HeartSine samaritan PAD (Public Access Defibrillator): Type/Catalog: SAM 350P/350-UNIT-US-10, 350-STR-US-10, 350-STR-UK-10; SAM 360P/360-STR-JA-10…

Why was HeartSine samaritan PAD recalled?

Automated external defibrillators were shipped in their test configuration so there is a potential to have incorrect language, and CPR duration resulting in the device prompting users to perform CPR for 5-seconds, and it could analyze heart rhythm and instruct users to deliver a shock every 5-seconds if a shockable heart rhythm is detected, potentially leading to no therapy or delayed therapy.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Type/UDI-DI or GTIN/Serial Number: SAM 350P/05060167120671, 05060167127670, +M727SAM350P0/$$+7/17D00023014, 18D00020508, 19D00017655, 20D00006524, 21D91081653, 22D91152391; SAM 360P/05060167127687, 05060167127311,5060167122927/21E91049810, 23E90001379, 23E90010409; SAM 450P/05060167129773, +M727SAM450P0/$$+7, +M727SAM450P0/$$+7/17G00001893, 18G00001553; SAM 500P/05060167122453/22B91071233

Product
HeartSine samaritan PAD (Public Access Defibrillator): Type/Catalog: SAM 350P/350-UNIT-US-10, 350-STR-US-10, 350-STR-UK-10; SAM 360P/360-STR-JA-10…
Recalling firm
HeartSine Technologies Ltd
Quantity
12
Distribution
Worldwide - US Nationwide distribution in the states of TX, CA, IL, CO, FL, NC, MI and the countries of Netherlands, Japan, New Zealand.
Recall date
February 21, 2024
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls