BD Vacutainer¿ EDTA Lavender Top Tubes recall
BD Vacutainer¿ EDTA Lavender Top Tubes: Catalog Numbers: 1. 366401 (UDI: 50382903664011); 2. 366450 (UDI: 50382903664509); 3. 366643 (UDI: 50382903666435); 4. 367835 (UDI: 50382903678353); 5. 367841 (UDI: 50382903678414); 6. 367844 (UDI: 50382903678445); 7. 367856 (UDI: 50382903678568); 8. 367861 (UDI: 50382903678612); 9. 367862 (UDI: 50382903678629); 10. 367863 (UDI: 50382903678636); 11. 368021 (UDI: 50382903680219); 12. 368047 (UDI: 50382903680479); 13. 368054 (UDI: 50382903680547); 14. 368661 (UDI: 50382903686617); 15. 364300 (UDI: pre-UDI)
Why was BD Vacutainer¿ EDTA Lavender Top Tubes recalled?
BD Vacutainer EDTA Lavender, Tan, and Pink Top Tubes and BD Vacutainer Lithium Heparin Green Top Tubes can cause an underestimation of lead in blood samples when used with Magellan Diagnostics LeadCare assays, employing the Anodic Stripping Voltammetry (ASV) methodology, or any other assay employing ASV methodology.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
All lots. Catalog Number (Shelf Life): 1. 366401 (24 months); 2. 366450 (24 months); 3. 366643 (12 months); 4. 367835 (16 months); 5. 367841 (15 months); 6. 367844 (16 months); 7. 367856 (16 months); 8. 367861 (16 months); 9. 367862 (16 months); 10. 367863 (17 months); 11. 368021 (16 months); 12. 368047 (15 months); 13. 368054 (16 months); 14. 368661 (17 months); 15. 364300 (Requested)
- Product
- BD Vacutainer¿ EDTA Lavender Top Tubes: Catalog Numbers: 1. 366401 (UDI: 50382903664011); 2. 366450 (UDI: 50382903664509); 3. 366643 (UDI: 50382903666435); 4. 367835 (UDI: 50382903678353); 5. 367841 (UDI: 50382903678414); 6. 367844 (UDI: 50382903678445); 7. 367856 (UDI: 50382903678568); 8. 367861 (UDI: 50382903678612); 9. 367862 (UDI: 50382903678629); 10. 367863 (UDI: 50382903678636); 11. 368021 (UDI: 50382903680219); 12. 368047 (UDI: 50382903680479); 13. 368054 (UDI: 50382903680547); 14. 368661 (UDI: 50382903686617); 15. 364300 (UDI: pre-UDI)
- Recalling firm
- Becton Dickinson & Company
- Quantity
- 1,116,012,050
- Distribution
- Distributed in 46 states: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NM, NY, NV, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and Puerto Rico.
- Recall date
- March 28, 2018
- Classification
- Class I
- Status
- Terminated
- Agency
- FDA (device)