SR Instruments FG3105 Patient Scales Used on various recall
SR Instruments FG3105 Patient Scales Used on various Joerns Healthcare Patient Lifters. Model# FG3105 Patient Scales. Product Usage: The FG3105 Scale…
Why was SR Instruments FG3105 Patient Scales Used on various recalled?
SR Instruments Inc. is implementing a field correction for the Joerns Healthcare Patient Lifts after they received two incident reports regarding two field failures of a tensile type load cell involving the FG3105 Patient Scale. The failures resulted in patient falls.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Serial numbers: 1005 thru 1780, 1785 thru 1809, 1866 thru 1868 and 1870 thru 1875.
- Product
- SR Instruments FG3105 Patient Scales Used on various Joerns Healthcare Patient Lifters. Model# FG3105 Patient Scales. Product Usage: The FG3105 Scale…
- Recalling firm
- Sr Instruments Inc
- Quantity
- 806 patient scales
- Distribution
- USA Nationwide Distribution in the state of Texas.
- Recall date
- April 24, 2013
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)