Stratum Lapidus Plate - Small Rt recall: Labeling Error
Stratum Lapidus Plate - Small Rt, Part No. STRM-LAP-SMR, UDI 00817701025467. The product is part of the Stratum Foot Plating System, a plate and…
Why was Stratum Lapidus Plate - Small Rt recalled?
The product contains a different plate than indicated by the package labeling. The product is labeled as a right plate but contains a left plate.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Lot M1502
- Product
- Stratum Lapidus Plate - Small Rt, Part No. STRM-LAP-SMR, UDI 00817701025467. The product is part of the Stratum Foot Plating System, a plate and…
- Recalling firm
- Nextremity Solutions
- Quantity
- 40
- Distribution
- Domestic distribution to AL, CA, FL, GA, ID, IL, IA, KY, LA, ME, MA, MI, MN, MS, NE, NJ, NY, OH ,OK, OR, TX, and WA.
- Recall date
- February 19, 2020
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)