FDAClass IIFebruary 19, 2020

SwishPlus Implant 5.7mmD x 8mmL recall

SwishPlus Implant 5.7mmD x 8mmL, SBM 6.5mmD Platform, Part Number 925708

Why was SwishPlus Implant 5.7mmD x 8mmL recalled?

Due to a complaint received and internal investigation it was determined that implant vials were distributed with incorrect color coded vials and caps. The part/lot numbers were correct, the customer received the product they ordered, however the color on the vial label and cap labels were incorrect.Specifically, the color coded vial and cap does not match the implant size described on the vial. As a result of the incorrect color on the vial and cap, the surgeon may not be certain if the implant received was the correct size. This could result in a delay in surgery, risk of infection if the surgical site has to be closed up and a second round of anesthesia.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot # 109038 No affected devices distributed within the US.

Product
SwishPlus Implant 5.7mmD x 8mmL, SBM 6.5mmD Platform, Part Number 925708
Recalling firm
Implant Direct Sybron Manufacturing LLC
Quantity
11 units
Distribution
US: None OUS: Iceland, Sweden, Spain,
Recall date
February 19, 2020
Classification
Class II
Status
Terminated
Agency
FDA (device)