FDAClass IIApril 17, 2019

COBAS INTEGRA 400 plus analyzer used in the following recall

COBAS INTEGRA 400 plus analyzer used in the following assays: a)Tina-quant Apolipoprotein A-1 ver.2 Catalog Number: 03032566122 b)Tina-quant Albumin…

Why was COBAS INTEGRA 400 plus analyzer used in the following recalled?

COBAS INTEGRA¿ 400 plus and cobas c 111 analyzers Additional Extra Wash Cycle Configurations and/or Batch Mode Needed for Specific Tests Due to Carry-over

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

All

Product
COBAS INTEGRA 400 plus analyzer used in the following assays: a)Tina-quant Apolipoprotein A-1 ver.2 Catalog Number: 03032566122 b)Tina-quant Albumin…
Recalling firm
Roche Diagnostics Corporation
Quantity
656 units
Distribution
US Nationwide distribution.
Recall date
April 17, 2019
Classification
Class II
Status
Terminated
Agency
FDA (device)