US Clinical recall
US Clinical 20mL Programmable Pump. Catalog No. 01827 The device is an implantable infusion pump. GTIN-14 00810335020082
Why was US Clinical recalled?
A pump alarm function anomaly in the pump firmware code may result in the pump not sending the expected error code notifications to the Programmer, or not emit an audible alarm tone in certain "out of tolerance" conditions associated with the pump's valve system.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
All units with pump firmware version 0.26 manufactured from 2012 until the present time. Serial Numbers 10HW4A57 10AX4A06 36ES4A15 36AR4A04 36AR4A06 36AR4A13 36HS4A10 36HS4A54 36JN4A04 36JN4A06 36JN4A07 36JN4A09 36JN4A13 36JN4A14 36JN4A15 36JN4A18 36JN4A23 36JN4A28 36JN4A35 36KN4A11 36KN4A16 36KN4A23 36KN4A27 36KN4A33 36KN4A34 36KN4A41 36HS4A01 10IT4A21 36HS4A14 36HS4A26 36BS4A17 36ES4A28 36ES4A33 36ES4A08 10IT4A16 36FR4A13 36HS4A61 10ET4A07 36HS4A46 10ET4A20 36HS4A25 36HS4A43 10ET4A02
- Product
- US Clinical 20mL Programmable Pump. Catalog No. 01827 The device is an implantable infusion pump. GTIN-14 00810335020082
- Recalling firm
- Flowonix Medical Inc
- Quantity
- 6670 (US)
- Distribution
- Nationwide domestic distribution. Foreign distribution to United Kingdom, Republic of Ireland, Poland, Greece, Italy, Germany, Netherlands, and Belgium.
- Recall date
- February 19, 2020
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)