FDAClass IIFebruary 8, 2017

Immulite/immulite 1000 Systems recall

IMMULITE/IMMULITE 1000 Systems, Third Generation PSA, Catalog Number/REF LKUP1, SMN 10380956 (OUS Kit Catalog) Product Usage: For in vitro diagnostic…

Why was Immulite/immulite 1000 Systems recalled?

Negative bias with IMMULITE¿ /IMMULITE¿ 1000 Third Generation PSA (prostate-specific antigen) kit. The investigation by Siemens has confirmed that patient samples exhibit an average negative bias of approximately -22% with a range of -16% to -27% for samples recovering from 0.08 to 16.2 ng/mL.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot Numbers: 1) 0336 - Expiration date 2017-01-31 2) 0337 - Expiration date 2017-02-28 3) 0338 - Expiration date 2017-05-31

Product
IMMULITE/IMMULITE 1000 Systems, Third Generation PSA, Catalog Number/REF LKUP1, SMN 10380956 (OUS Kit Catalog) Product Usage: For in vitro diagnostic…
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
1379 kits (190 kits in U.S.A and 1189 Kits internationally) in total
Distribution
Worldwide Distribution - US Nationwide and the countries of Denmark, Brazil, Croatia and Germany
Recall date
February 8, 2017
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls