FDAClass IIApril 17, 2019

SORIN Heater-Cooler System 3T recall

SORIN Heater-Cooler System 3T, REF 16-02-85 Intended for use with a Stockert/Sorin S3/S5/C5 heart-lung machine and/or any other heart-lung machine…

Why was SORIN Heater-Cooler System 3T recalled?

For six units, the post-deep cleaning (DC) water sample tested positive for nontuberculous mycobacteria (NTM).

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Serial numbers 16S13187, 16S14887, 16S15335, 16S15336, 16S15337, and 16S11877.

Product
SORIN Heater-Cooler System 3T, REF 16-02-85 Intended for use with a Stockert/Sorin S3/S5/C5 heart-lung machine and/or any other heart-lung machine…
Recalling firm
LivaNova USA Inc.
Quantity
6 devices
Distribution
US Distribution to states of: IN, MO, and OH.
Recall date
April 17, 2019
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls