FDAClass IIIMay 25, 2022

CK MB Calibrator recall

CK MB Calibrator, Catalogue Number CK2393

Why was CK MB Calibrator recalled?

Randox have realigned CK-MB Calibrator, CK2393, lot number 4410CK. Calibrating with the new value will cause a positive shift of approximately 6% with both Quality Controls and Patient Samples.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

GTIN 05055273201451 Lot 4410CK

Product
CK MB Calibrator, Catalogue Number CK2393
Recalling firm
Randox Laboratories Ltd.
Quantity
1,392
Distribution
Domestic distribution to WV and PR. International distribution to Bangladesh, Bosnia & Herzegovina, Botswana, China, Colombia, Costa Rica, Croatia, Egypt, Fiji, France, Georgia, Germany, Ghana, Greece, Grenada, Hong Kong, India, Iraq, Italy, Lithuania, Malaysia, Maldives, Malta, Mexico, Morocco, Mozambique, Nepal, Nigeria, Pakistan, Philippines, Poland, Puerto Rico, Romania, Russia, Saudi Arabia, South Korea, Switzerland, Thailand, Turkey, UK, Uruguay, Uzbekistan and Vietnam.
Recall date
May 25, 2022
Classification
Class III
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls