FDAClass IIFebruary 12, 2020

SafeAir Telescopic Smoke Evacuation Pencil recall

SafeAir Telescopic Smoke Evacuation Pencil, Product No. 0703 046 004. The product is a Bovie pencil that is 75 inches long compressed, with a…

Why was SafeAir Telescopic Smoke Evacuation Pencil recalled?

Hairline fractures in a component could allow electrical current to arc out of the device, resulting in a burn to the patient and/or user.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

1921032

Product
SafeAir Telescopic Smoke Evacuation Pencil, Product No. 0703 046 004. The product is a Bovie pencil that is 75 inches long compressed, with a…
Recalling firm
Stryker Instruments Div. of Stryker Corporation
Quantity
15150
Distribution
Domestic distribution nationwide. Foreign distribution to Netherlands and Australia.
Recall date
February 12, 2020
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls